Fume cupboards are the front line of containment in any laboratory or workshop handling volatile, toxic or sensitising substances. COSHH 2002 requires them to be tested and examined at suitable intervals by a competent person — but the specifics are easy to lose in the noise. This article sets out what fume cupboard testing actually involves, the standards that govern it, and how to spot a unit that is failing quietly.
Regulation 9 of COSHH classifies fume cupboards as local exhaust ventilation (LEV) and brings them under the 14-month thorough examination requirement. The test verifies that the cupboard is still containing as designed — not just running.
The European standard for fume cupboard performance, covering type-testing (manufacturer), on-site testing (commissioning) and in-service testing (routine). Most UK in-service tests reference Part 4 of the standard.
HSE guidance on the design, testing and management of LEV. The reference text for competent persons doing in-service testing across all LEV types, including fume cupboards.
At minimum every 14 months under COSHH. Most labs run a 12-month cycle. For cupboards in heavy use or handling high-hazard substances, six-monthly testing is sensible — and may be specified by your insurer.
Daily and weekly checks by users matter as much as the annual test. Users should:
Every fume cupboard test produces a report. Keep it for at least five years. The HSE will ask for it in any LEV inspection. Build a simple register: cupboard ID, location, last test date, next test due, airflow values and any defects raised.
A competent fume cupboard tester holds a P601 qualification (or equivalent) and has documented experience with the specific cupboard type. ESiNet matches you with an engineer whose accreditation and lab-sector experience fit your environment.
Request an inspection of your fume cupboards or wider LEV system.
No. Once it has failed containment, it is out of service until repaired and re-tested. Continuing to use it puts users at risk and creates direct COSHH exposure.
Same COSHH framework, different test methodology. The competent person will adapt the protocol to suit the geometry, sash arrangement and use pattern.
Only for the substances the filter is rated for, and only if filter saturation is monitored. Many sites use them outside their rating, which is a common LEV finding.
Same picture as recirculating units. Filter type, substance compatibility and saturation monitoring are the make-or-break checks.